IEEE 488.2 and Specific Command Reference
418 FTB-500
Specific Commands
Response(s) Hour:
The response data syntax for <Hour> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Hour> ranges from 0 to 23. The
instruments returns hour information in 24-hour
format.
Minute:
The response data syntax for <Minute> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Minute> ranges from 0 to 59.
Second:
The response data syntax for <Second> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Second> ranges from 0 to 59. The
resolution of the clock is the second.
Example(s) SYST:TIME? returns 16,55,38
See Also SYSTem:TIME