Справка по командам SCPI
438 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
WIDTh[:WAVelength]?
Description This query returns the wavelength width of the
WDM analysis default channel.
At *RST, default channel width is set to 50.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:WIDTh[
:WAVelength]?[<wsp>MAXimum|MINimum|DEFa
ult]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Width>