SCPI Command Reference
Optical Spectrum Analyzer 471
Product-Specific Commands—Description
:CALCulate[1..n]:RBW:INSTrument
Description This command sets whether the application uses
the Instrument RBW or the user defined RBW
value.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 1 (or ON).
Syntax :CALCulate[1..n]:RBW:INSTrument<wsp><Rb
wInstrumentState>
Parameter(s) RbwInstrumentState:
The program data syntax for
<RbwInstrumentState> is defined as a
<Boolean Program Data> element. The
<RbwInstrumentState> special forms ON and
OFF are accepted on input for increased
readability. ON corresponds to 1 and OFF
corresponds to 0.
The <RbwInstrumentState> special forms ON
and OFF are accepted on input for increased
readability. ON corresponds to 1 and OFF
corresponds to 0.