Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Bedienungsanleitung Seite 526

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 653
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 525
SCPI-Befehlsreferenz
518 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]
Description This command sets custom resolution
bandwidth value for WDM analysis OSNR
calculation.
At *RST, this value is set to 0.100 nm.
Syntax :CALCulate[1..n][:WDM]:OSNR:BANDwidth|BWI
Dth[:RESolution]<wsp><Resolution[<wsp>M]
>|MAXimum|MINimum|DEFault
Parameter(s) Resolution:
The program data syntax for <Resolution> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is M. The
<Resolution> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Seitenansicht 525
1 2 ... 521 522 523 524 525 526 527 528 529 530 531 ... 652 653

Kommentare zu diesen Handbüchern

Keine Kommentare