SCPI Command Reference
OSA 605
Examples on Using the SCPI Commands
Creating a Channel List Based on Specific Channels
(WDM)
Command Sequence Comments
CALC:WDM:CHAN:AUTO OFF Disable default channel.
CALC:WDM:CHAN:DEL:ALL Clear current channel list.
CALC:WDM:CHAN:DEF
"CWDM_1470",1470.0 NM
Add a new channel named "CWDM_1470" with
nominal central wavelength at 1470.0 nm.
All others parameters for this new channel are
set to their default values.
CALC:WDM:CHAN:SEL "CWDM_1470" Select channel "CWDM_1470".
<Add commands to modify channel
parameters>
See Modifying Selected Channel Analysis
Parameters (WDM) on page 608.
CALC:WDM:CHAN:DEF
"CWDM_1530",1530.0 NM
Add a new channel named "CWDM_1530" with
a nominal central wavelength at 1530.0 nm.
All others parameters for this new channel are
set to their default value.
CALC:WDM:CHAN:SEL "CWDM_1530" Select channel "CWDM_1530".
<Add commands to modify channel
parameters>
See Modifying Selected Channel Analysis
Parameters (WDM) on page 608.
CALC:WDM:CHAN:DEF
"CWDM_1550",1550.0 NM
Add a new channel named "CWDM_1550" with
a nominal central wavelength at 1550.0 nm.
All others parameters for this new channel are
set to their default value.
CALC:WDM:CHAN:SEL "CWDM_1550" Select channel "CWDM_1550".
<Add commands to modify channel
parameters>
See Modifying Selected Channel Analysis
Parameters (WDM) on page 608.