Remote Control (SCPI) Commands
High-Speed Power Meter 121
Specific Commands
INITiate:CONTinuous
Description This command starts or stops power measurements (for both
channels when using a PM-1620) in power measurement
mode.
Syntax INIT:CONT<space><boolean>
Parameters The <boolean> value refers to
“0”- stop measurements
“1”- start measurements
Example INIT:CONT 1
See also FETC[:SCAL]:POW:DC?, INIT:CONT?, INIT[:IMM], and
READ[:SCAL]:POW:DC?
INITiate:CONTinuous?
Description This query returns a value indicating whether power
measurements are in progress (for bothchannels when using a
PM-1620) in power measurement mode.
Syntax INIT:CONT?
Response “0”- stopped
“1”- started
Example INIT:CONT?
See also FETC[:SCAL]:POW:DC?, INIT[:IMM], and
READ[:SCAL]:POW:DC?
Remote Control (SCPI) Commands
High-Speed Power Meter 121
Specific Commands
INITiate:CONTinuous
Description This command starts or stops power measurements (for both
channels when using a PM-1620) in power measurement
mode.
Syntax INIT:CONT<space><boolean>
Parameters The <boolean> value refers to
“0”- stop measurements
“1”- start measurements
Example INIT:CONT 1
See also FETC[:SCAL]:POW:DC?, INIT:CONT?, INIT[:IMM], and
READ[:SCAL]:POW:DC?
INITiate:CONTinuous?
Description This query returns a value indicating whether power
measurements are in progress (for bothchannels when using a
PM-1620) in power measurement mode.
Syntax INIT:CONT?
Response “0”- stopped
“1”- started
Example INIT:CONT?
See also FETC[:SCAL]:POW:DC?, INIT[:IMM], and
READ[:SCAL]:POW:DC?