Referencia de instrucciones SCPI
520 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:DATA:
SIGnalPower:FLATness?
Description This query returns computed WDM analysis
global result for signal power flatness.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:DATA:SIGnalPower:FLA
Tness?
Parameter(s) None
Response Syntax <Flatness>
Response(s) Flatness:
The response data syntax for <Flatness> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Flatness> response corresponds to the
computed signal power flatness.
Example(s) <Do measurement>
UNIT:RAT DB
CALC:DATA:SIGP:FLAT? Returns 3.118000E+001
Notes Special NAN (not a number) value
-2251799813685248 is returned if analysis result
could not be computed.
See Also :CALCulate[1..n][:WDM]:DATA:OSNR:FLATness?
:CALCulate[1..n][:WDM]:DATA:OSNR:MEAN?
:CALCulate[1..n][:WDM]:DATA:SIGnalPower:ME
AN?