Referencia de instrucciones SCPI
566 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:TYPE?
Description This query returns the selected averaging type for
trace acquisition.
At *RST, averaging is set to SCALar.
Syntax :SENSe[1..n]:AVERage:TYPE?
Parameter(s) None
Response Syntax <Type>
Response(s) Type:
The response data syntax for <Type> is defined
as a <CHARACTER RESPONSE DATA> element.
The <Type> response corresponds to the
selected averaging type.
SCALAR: scalar averaging type is selected.
POLARIZATIONMINMAXHOLD: averaging type
for InBand noise measurement is selected.
Example(s) SENS:AVER:TYPE PMMH
SENS:AVER:TYPE?
POLARIZATIONMINMAXHOLD
See Also :SENSe[1..n]:AVERage:COUNt
:SENSe[1..n]:AVERage:TYPE