SCPI Command Reference
OTDR 443
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:POWer:STATe:
TIME
Description This command sets the duration after which the
visual fault locator (VFL) will stop emitting light
automatically (auto-off feature).
Note that this command does not turn the VFL
on.
*RST sets this value to 600 seconds.
Syntax :SOURce[1..n]:VFLocator:POWer:STATe:TIME<
wsp><Duration>|MAXimum|MINimum|DEFau
lt
Parameter(s) Duration:
The program data syntax for <Duration> is
defined as a <numeric_value> element. The
<Duration> special forms MINimum, MAXimum
and DEFault are accepted on input.