Exfo FTB-7000 OTDR Series for FTB-500 Bedienungsanleitung Seite 422

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 493
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 421
SCPI-Befehlsreferenz
408 Serie FTB-7000
Product-Specific Commands—Description
:FETCh[1..n]:ASETting:DURation?
Description This query returns the duration found after an
initiate (INIT) command. Note that acquisition
mode (CONF:ACQ:MODE) must be set to
ASETting.
Since *RST clears the duration value, the
returned value will be 0.
Syntax :FETCh[1..n]:ASETting:DURation?
Parameter(s) None
Response Syntax <Duration>
Response(s) Duration:
The response data syntax for <Duration> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
Returns the duration, in seconds.
Example(s) CONF:ACQ:MODE ASET
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
FETC:ASET:DUR? Ex.: Returns 15
See Also CONFigure[1..n]:ACQuisition:DURation?
Seitenansicht 421
1 2 ... 417 418 419 420 421 422 423 424 425 426 427 ... 492 493

Kommentare zu diesen Handbüchern

Keine Kommentare