Exfo FTB-7000 OTDR Series for FTB-500 Bedienungsanleitung Seite 463

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 493
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 462
SCPI-Befehlsreferenz
OTDR 449
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:AM:INTernal:
FREQuency?
Description This query returns a value indicating the current
internal modulation frequency. If the visual fault
locator (VFL) is in CW mode, the function will
return 0.
*RST sets the modulation frequency to 0 Hz
(CW).
Syntax :SOURce[1..n]:VFLocator:AM:INTernal:FREQuen
cy?[<wsp>MINimum|MAXimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Frequency>
Seitenansicht 462
1 2 ... 458 459 460 461 462 463 464 465 466 467 468 ... 492 493

Kommentare zu diesen Handbüchern

Keine Kommentare