SmartR™ Features
138 MAX-635
FaultMapper
FaultMapper Result Details
The Result Details read-only page displays the test
results in text and graphical format.
Test Identification is the name of the test, for
example FaultMapper.
Cable Setup button opens a new page displaying
the cable parameters of the test. Use the left/right
arrows to select the button and press to open
the page.
Result displays the cable fault details.
Note: Some displayed resistive faults may be effective faults.
For example if a strap is present, a T-G short may be shown twice: first as a
T-G short and second as an effective R-G short. In such cases, the
resistance shown is the resistance that would be measured using a
multimeter.
Estimated Length displays the cable length when the cable is open.
Fault Status displays the number of faults detected (or No fault