Remote Control Commands
Optical Power Meter 81
Product-Specific Commands—Description
INITiate[:IMMediate]
Description This command stores a value in the initialized memory location
so it can be fetched at any time with the FETC[:SCAL]:POW:DC?
query. Values can be power or loss measurements, depending
on the current measurement mode of the PM-1100.
Syntax INIT[:IMM]
Parameters N/A
Example INIT:IMM
Note Use the command INIT:STOR to initialize a memory location.
See also FETC:SCAL:POW:DC?, INIT:CONT, READ:SCAL:POW:DC?
INITiate:STORage
Description This command allows you to initialize a memory location (out
of the 512 available) before storing or recalling an absolute or
referenced measurement.
Syntax INIT:STOR<space><data>
Parameters The <data> parameter is a memory location. This value can be
any value between 1 and 512, inclusive.
Example INIT:STOR 99
Note Use the INIT:CONT or INIT:IMM commands to store a value in
the initialized memory location.
Use the FETC[:SCAL]:POW:DC command to fetch the value
currently in the initialized memory location.
See also INIT:IMM, INIT:CONT, and FETC[:SCAL]:POW:DC