Remote Control Commands
82 PM-1100
Product-Specific Commands—Description
MMEMory:ACQuisition
Description This command initiates a data acquisition with the currently set
parameters (delay, duration, number of samples). The
acquisition will continue until the duration elapses or until an
acquisition stop (MMEM:ACQ 0) command is sent.
Syntax MMEM:ACQ<space><boolean>
Parameters The <boolean> parameter can be
“1” start the data acquisition
“0” stop the data acquisition
Example MMEM:ACQ 1
MMEMory:ACQuisition?
Description This query returns a value indicating whether data acquisition is
in progress.
Syntax MMEM:ACQ?
Response The response is in the format
“1” data acquisition is in progress
“0” data acquisition is not in progress
Example MMEM:ACQ?