Remote Control
118 FLS-2600B
Programming Commands
Programming Commands
SOURce:SWEep:CONTinuous:PROGram
Description This function is used to set the parameters for a continuous
sweep program. During a sweep program, the application
sweeps continuously between the minimum (start) and
maximum (stop) wavelengths.
Syntax SOUR:SWE:CONT:PROG<wsp>[<start>],[<stop>],
[<speed>],[<coun>],[<rev>],[<trig>]
Parameters The <start> and <stop> parameters represent the start and stop
wavelengths for the sweep. The current spectral measurement
unit applies (nm or THz). The format must be 9999.999 nm or
999.9999 THz.
The <speed> parameter is used to set the speed of the sweep.
The format must be 99.999.
The <coun> parameter is the number of passes. The format
must be an integer from 1 to 100 or 0 for continuous.
The <rev> parameter indicates whether the reverse option is on
or off. The format must be 1|0.
The <trig> parameter indicates whether the trigger option is on
or off. The format must be 1|0.