Exfo FLS-2600B Tunable Laser Source Bedienungsanleitung Seite 95

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 157
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 94
Remote Control
Tunable Laser Source 87
General Commands
*RST
Description The *RST command sets the instrument to reset settings
(factory settings stored in ROM). Pending *OPC and *OPC?
actions are canceled. The instrument becomes idle. The *RST
command clears the Input and Error queues. The following
settings remain unchanged:
Instrument interface address
Output queue
Service Request Enable register (SRE)
Standard Event Status Enable register (ESE)
Syntax *RST
Note For a complete list of default parameters, see Saving and
Recalling Configurations on page 25.
This command uses electronic parts that have a limited life
span (non-volatile memory). You must use this command with
moderation.
Seitenansicht 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 156 157

Kommentare zu diesen Handbüchern

Keine Kommentare