Remote Control
Tunable Laser Source 137
FLS-2600B Error and Warning Messages
FLS-2600B Error and Warning Messages
Error or
warning
Number
Description Recommended Action
11 Nulling was not performed
correctly.
Restart your Tunable Laser
Source.
12 The Tunable Laser Source
returns the wrong
identification.
Contact EXFO.
20 There was a communication
error with the module.
Restart your Tunable Laser
Source.
25 There was a checksum error
while reading the source’s
FIFO.
Restart your Tunable Laser
Source.
30 The command was not
accepted and has been lost.
Verify if program resumed after
having lost a command. Rerun
the program.
31 A setting sent to the tunable
laser source is wrong.
Review command before
sending it again.
32 Command sent while
tunable laser source is
active.
Wait until Tunable Laser Source
is done before sending a
command.
34 Too many commands sent at
the same time.
Wait until the Tunable Laser
Source is done before sending
more commands.
40 The FIFO in the Tunable
Laser Source is not ready for
reading.
Wait for the FIFO to be ready.
60 A problem occurred with the
current going to the laser.
Contact EXFO.
Seitenansicht 144