SCPI Command Reference
104 IQS-3150
Product-Specific Commands—Description
STAT:OPER:BIT8:COND? Keep sending the query
as long as attenuation is not reached. When
attenuation is reached, the returned value is 0.
INP:ATT? Returns: 1.435500E+001
INP:RATT? Returns: 1.535500E+001
OUTP:APM REF
INP:ATT? Returns: 1.435500E+001
INP:RATT? Returns: 1.000000E+000
INP:RATT -2.000
STAT:OPER:BIT8:COND? Keep sending the query
as long as attenuation is not reached. When
attenuation is reached, the returned value is 0.
INP:ATT? Returns: 1.135500E+001
INP:RATT? Returns: -2.000000E+000
INP:RATT MIN
See Also :CONTrol[1..n]:MODE
:OUTPut[1..n]:APMode
:INPut[1..n]:ATTenuation
:INPut[1..n]:OFFSet
:INPut[1..n]:RATTenuation?
:INPut[1..n]:REFerence
:STATus:OPERation:BIT<n>:CONDition?
:OUTPut[1..n]:RPOWer
:INPut[1..n]:RATTenuation