SCPI Command Reference
144 IQS-3150
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:COLLect:
ZERO
Description This command performs an offset nulling on the
internal power meter.
This command is an event and has no associated
*RST condition or query form.
Syntax :SENSe[1..n]:CORRection:COLLect:ZERO
Parameter(s) None
Example(s) STAT? Must return READY
SENS:CORR:COLL:ZERO
STAT:OPER:BIT10:COND? Keep resending the
query as long as the operation is not complete
(returned value is not 0).
Notes This command will take at least 3 seconds to
complete.
See Also :CALibration[1..n]:ZERO
:STATus:OPERation:BIT<n>:CONDition?