SCPI Command Reference
112 IQS-3150
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Wavelength> parameter.
The <Wavelength> parameter corresponds to
the current wavelength. You can use the
INPut[1..n]:WAVelength? MAX and
INPut[1..n]:WAVelength? MIN queries to
determine a valid range for the wavelength.
Example(s) INP:WAV 1310 NM
STAT:OPER:BIT8:COND? Keep sending the query
as long as attenuation is not reached. When
attenuation is reached, the returned value is 0.
INP:WAV 0.000001550 M
STAT:OPER:BIT8:COND? Keep sending the query
as long as attenuation is not reached. When
attenuation is reached, the returned value is 0.
See Also :INPut[1..n]:WAVelength?
:STAT:OPER:BIT:COND?
:INPut[1..n]:WAVelength