SCPI Commands Reference
DWDM Passive Component Test System 249
Specific Commands
REFerence:IL
Description This command performs an IL reference acquisition.
Syntax REF:IL
Parameter None
Response None
Example REF:IL
REFerence:IL?
Description The response to this query indicates if an IL reference
acquisition has been performed.
Syntax REF:IL?
Parameter None
Response A boolean response representing the state of the IL reference
acquisition:
“1” - The IL reference acquisition is performed
“0” - The IL reference acquisition is not performed
Example REF:IL?