COM Objects Reference
DWDM Passive Component Test System 275
Measures Interface—Methods
Measures
AcqIlPdlDut
Description This command performs an IL/PDL measurement on the DUT.
Syntax object.AcqllPdlDut(FireEvent)
Parameters Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrHardNotCompatible, ErrCalWrBefore,
ErrRefIlPdlBefore, ErrInProgress, ErrAcqType, ErrAcqStopped,
ErrScanRange, ErrSourceNotEnabled. For more information,
see IQS-12004B DWDM Passive Component Test System
(Standard) COM Error Messages on page 208.
Note Available only with the IQS-5150 polarization state adjuster.
Measures
AcqOrlMandrelDut
Description This command performs an ORL measurement on the DUT.
Syntax object.AcqOrlMandrelDut(FireEvent)
Parameters Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrCalOrlRtjBefore, ErrCalOrlMandrelRtjBefore,
ErrRefOrlMandrelBefore, ErrInProgress, ErrAcqType,
ErrAcqStopped, ErrScanRange, ErrSourceNotEnabled. For more
information, see IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages on page 208.
Kommentare zu diesen Handbüchern