Setting Up the DWDM Passive Component Test System
56 IQS-12004B
Modifying System Settings
Modifying System Settings
Note: You must be working at the Supervisor level to use this feature.
The System Settings window allows you to define two saving options, set
filtering options, and select a new database folder. To access this window,
from the main window menu, select Options > Supervisor Options >
System Settings.
³ Save Raw Data: This is a troubleshooting tool that allows EXFO
technical support to troubleshoot data analysis problem. This option is
normally not selected as it uses a large storage space.
³ Save Results Curves (IL, PDL, ORL): This option is selected by
default. If not selected, you will only be able to see the results table
(not the trace) and the trace data will not be saved.