SCPI Commands Reference
DWDM Passive Component Test System 261
Specific Commands
READ:WavelengthReferenceModule?
Description The response to this query indicates the power of both
channels on the wavelength reference module.
Syntax READ:WRM?
Parameter None
Response A string in the “Reference channel, ORL channel” dBm format,
i.e., “9999.999,9999.999” dBm format.
Example READ:WRM?
Note This is a blocking function i.e., In progress appears in the
status bar and you must wait before sending a new command.
READ:SerialNumber?
Description The response to this query indicates the serial number of each
modules initialized in the system.
Syntax READ:SN?
Parameter None
Response A string in the “ModuleName SlotPosition: Serial Number”
format.
Example READ:SN?
Note Each module description is separated by a comma.