Exfo IQS-12004B DWDM Passive Component Test System for Bedienungsanleitung Seite 273

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 437
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 272
SCPI Commands Reference
DWDM Passive Component Test System 261
Specific Commands
READ:WavelengthReferenceModule?
Description The response to this query indicates the power of both
channels on the wavelength reference module.
Syntax READ:WRM?
Parameter None
Response A string in the “Reference channel, ORL channel” dBm format,
i.e., “9999.999,9999.999” dBm format.
Example READ:WRM?
Note This is a blocking function i.e., In progress appears in the
status bar and you must wait before sending a new command.
READ:SerialNumber?
Description The response to this query indicates the serial number of each
modules initialized in the system.
Syntax READ:SN?
Parameter None
Response A string in the “ModuleName SlotPosition: Serial Number
format.
Example READ:SN?
Note Each module description is separated by a comma.
Seitenansicht 272
1 2 ... 268 269 270 271 272 273 274 275 276 277 278 ... 436 437

Kommentare zu diesen Handbüchern