Exfo IQS-12004B DWDM Passive Component Test System for Bedienungsanleitung Seite 407

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 437
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 406
COM Objects Reference
DWDM Passive Component Test System 395
NotchResult—Properties
NotchResult
BwUpperBound
Description This read-only property returns the upper-bound wavelength
value over the bandwidth interval, based on the current power
level (from the peak). The power level is stated by the
PowerLevel property of the parent object. BwUpperBound is
expressed in nm.
Syntax object.BwUpperBound
Parameters None
Response A Double value
Access Get
NotchResult
CentralWavelength
Description This read-only property returns the wavelength at the midpoint
between the lower and the upper wavelengths based on the
power level of the parent object. The value is expressed in nm.
Syntax object.CentralWavelength
Parameters None
Response A Double value
Note The midpoint that is returned is calculated for the current
power level (from the peak) as specified in the PowerLevel
property of the parent object.
Access Get
Seitenansicht 406
1 2 ... 402 403 404 405 406 407 408 409 410 411 412 ... 436 437

Kommentare zu diesen Handbüchern

Keine Kommentare